B Braun Infusomat-Blood Administration
Clinical Practice Issue
Reports of frequent Air in Line (AIL) alarms during blood administration with B Braun Infusomat Pump that interferes with timely blood administration to patient.
Solution to address issue
Use of the Infusomat Space Pump Y-Type Blood Set- Ref 363419 (SH-SS SKU 91358) with Anti-Siphon Valve (ASV) left on has decreased the frequency of AIL alarms for blood administration. Product is available to order through SH-SS Logistics and Supply Chain Management.

Clinical consideration
Direction to use ASV is specific to this IV set. Be aware that the ASV on IV tubing is not recommended for general use because it interferes with bolus administration of IV fluids. For complete details see: Shared Health Safer Practice Notice-Discontinuing the Use of the Anti-Siphon Valve
For printable version.